What Affects Inward Leakage and Total Inward Leakage Results in RPE Testing?

What affects inward leakage and total inward leakage results, EN 13274-1 testing at GCS

Why is my respirator leaking during testing?

Inward leakage is the result manufacturers question most during development and approval testing. Why is it leaking, where is the leak coming from, and what is causing it? There is rarely a single answer.

The leakage measured in a test is the combined result of the device’s design and components, how it is fitted, how the wearer moves and breathes, and how the test is set up. Sometimes one factor accounts for most of the result. Often several combine.

This article explains some of the factors that can influence inward leakage (IL) and total inward leakage (TIL) results when respiratory protective devices (RPD) are tested to EN 13274-1:2001. It covers filtering facepieces, half masks, full face masks, powered air-purifying respirators (PAPRs), powered hoods and airline devices, and it focuses on laboratory testing rather than workplace use.

 

Inward leakage vs total inward leakage: the EN 13274-1 definitions

EN 13274-1 sets out three related measurements:

  • Inward leakage of a facepiece. Leakage into a facepiece, such as an EN 136 full face mask or EN 140 half mask, assessed on its own.
  • Inward leakage of a device (IL). The total inward leakage excluding any filter penetration. It covers every route into the breathing zone except through the filter media: the face seal, valves, diaphragms, seals, joints and connections.
  • Total inward leakage (TIL). Leakage from all sources, including penetration through the filter.

In each case, leakage is expressed as the concentration of test agent measured inside the facepiece as a percentage of the challenge concentration in the test enclosure.

To measure IL on a filtering device, the filter’s contribution has to be taken out. The device is supplied with clean air instead, with its breathing resistance kept similar to the complete device. Alternatively, gas or combined filters can be replaced by high-efficiency particle filters of the same mass and resistance. Facepieces with an EN 148-1 threaded connection are tested with a filter simulator fitted and clean air supplied through a lightweight hose.

EN 13274-1 is a test method, not a product standard. The relevant device standard decides which measurement applies, which method is used, how many samples and test subjects are needed, and any conditioning before the test.

The three EN 13274-1 measurements: facepiece inward leakage, device inward leakage and total inward leakage

How inward leakage testing works

A test subject wearing the device walks on a treadmill at 6 km/h inside an enclosure. A challenge atmosphere flows down over their head. A probe sealed through the facepiece, as close to the mouth as practical, samples the air inside. A second probe samples the enclosure.

The background level inside the facepiece is measured first with the challenge atmosphere off. Once the atmosphere is on and stable, the subject performs five exercises of two minutes each while walking:

  • Walking without head movement or talking.
  • Turning the head from side to side.
  • Moving the head up and down.
  • Reading the alphabet or an agreed text aloud.
  • Walking without head movement or talking.

Where the device standard requires it, some exercises are repeated with supplementary fans producing a 2 m/s crosswind from the front, side and rear. Leakage is calculated from the last 100 seconds of each exercise, so one exercise does not carry over into the next.

 

Choosing the test agent

EN 13274-1 uses two challenge agents:

  • Sodium chloride (NaCl) aerosol. A polydisperse salt aerosol with a mean enclosure concentration of 8 ± 4 mg/m³, measured with a flame photometer.
  • Sulfur hexafluoride (SF₆) gas. A tracer gas at around 0.1% to 1% by volume, measured with a gas analyser.

The choice depends on the device type and whether its materials are permeable (Table 1 of the standard). SF₆ is not retained by filters, so it suits IL measurement but not TIL of particle filtering devices. NaCl is not retained by gas filters. Where facepiece or hood materials are permeable to air, SF₆ is used.

With NaCl, the inside of the facepiece is sampled in one of two ways. Pulsed sampling (method 2A) only samples during inhalation. Continuous sampling (method 2B) samples throughout the breathing cycle and applies a correction for salt retained in the lungs. The device standard specifies which method applies.

How an EN 13274-1 inward leakage test runs: five two-minute exercises on a treadmill at 6 km per hour

Device design and components

Any route into the breathing zone that bypasses the filter counts towards inward leakage. On most devices there are several such routes, and each behaves differently as the wearer moves, speaks and breathes.

Component Typical devices How it can affect leakage
Face seal Filtering facepieces, half masks, full face masks Seal shape, material and flexibility decide how well the edge follows the face, including as the jaw and head move.
Exhalation valve Filtering facepieces, half masks, full face masks The valve must close fully on inhalation. Disc thickness and stiffness, seat geometry and how the disc sits all affect this.
Inhalation valve Half masks, full face masks Gaps around the valve housing or its seat in the facepiece can let air bypass the filter.
Speech diaphragm Full face masks, some half masks The joint between the diaphragm, its clamping ring and the housing is a potential leakage path.
Visor-to-body seal Full face masks, PAPR and airline headtops The seal between the visor and the mask body or frame must hold under head movement and pressure changes.
Neck seals and bibs Hoods, loose-fitting PAPRs and airline devices Without a tight face seal, the neck seal or bib limits how easily air can enter around the neck and chin.
Filter connections Filter-equipped masks and PAPRs Threads, bayonets and gaskets must seal; air leaking around the filter is not filtered.
Hose and other joints PAPRs, airline devices, masks with adapters Every connection, seam or bonded joint is a possible path.
Head harness Most tight-fitting devices Strap tension and adjustment range decide whether the device stays in place during head movement.

Component fit matters as much as design. Parts that do not seat consistently, or vary within their tolerances, can produce different results on different samples of the same design. This is why device standards test several samples, not just one.

 

Flow rate, filters and materials

Air flow rate in powered and airline devices

For PAPRs and powered hoods (EN 12941 and EN 12942) and compressed airline devices (EN 14594), the supplied air flow is central to the leakage result. These devices are typically tested at the manufacturer’s minimum design flow rate.

The supplied air keeps the inside of the headtop or facepiece at a higher pressure than the surroundings. If the wearer’s inhalation flow at any moment exceeds the supplied flow, the pressure inside can briefly drop below ambient. Air can then be drawn in through any gap. A lower supply flow leaves less margin for these peaks, which is why a low flow can lead to higher inward leakage.

On loose-fitting devices such as hoods, this airflow, together with the neck seal or bib, is what limits ingress. How the air is distributed inside the headtop can also affect how well the breathing zone stays clear of the challenge agent.

 

Filter penetration (TIL only)

TIL includes the challenge aerosol that passes through the filter as well as leakage through other routes. The filter’s efficiency therefore contributes directly to the TIL result. IL excludes it.

 

Material permeability and seams

Some hoods and headtops use fabrics or films that air can pass through. EN 13274-1 includes a bubble test to check this. A wetted material sample is pressurised up to 100 mbar, and continuous bubbles show it is permeable. Samples are taken from different areas, including seams. Where materials are permeable, SF₆ is used, and gas passing through the material or seams is measured as leakage.

 

Fit, the test panel and the test set-up

Fitting instructions

Test subjects read the manufacturer’s fitting instructions and are shown how to fit the device if needed. The fit is checked visually before the test starts, and the device is fitted as the instructions describe. The instructions therefore shape the result. Unclear, ambiguous or incomplete instructions, such as on strap order, nose clip adjustment or how to position a bib, can lead to inconsistent fitting across subjects.

Subjects may adjust the device during the test. If they do, that section of the test is repeated once readings have settled.

 

Test subjects and sizing

Test subjects are people familiar with using similar equipment. Together they cover a spread of facial characteristics, and four facial dimensions are recorded for each: face length, face width, face depth and mouth width. Where the manufacturer states a face or head size range, only subjects within it are used. Where several sizes are made, subjects choose the most suitable one using the manufacturer’s information.

Each subject is asked whether the device fits. If a subject says it does not, they are removed from the panel and this is reported. The number of subjects and how they are selected are set by the device standard. The sizes offered and the sizing guidance can therefore affect results across the panel.

 

Exercises and breathing

Each exercise loads the device differently. Head turns and up-and-down movements shift the harness, hoses and headtop. Speaking moves the jaw against the seal. Walking at 6 km/h raises breathing rate and peak inhalation flow. Breathing is not controlled, so it varies naturally between subjects. Subjects receive no indication of their results while the test runs.

 

Preconditioning

Some device standards require samples to be conditioned before leakage testing, for example by temperature exposure. Changes in materials, valves or seals after conditioning can show up in the leakage result.

 

The test set-up

The set-up itself has to avoid influencing the result, and EN 13274-1 sets requirements for this:

  • The sampling probe must be fitted airtight through the facepiece, as near the mouth as possible.
  • On flexible or soft facepieces, a separate head harness may carry the probe so its weight does not distort the device.
  • Clean-air hoses, filter simulators and surrogate filters must not affect the fit or replace any of the device’s own seals.
  • Devices with small internal volumes or soft hoods can make probe placement more difficult. Every measurement also carries an estimate of uncertainty, which is stated in the test report under Annex A of the standard.

Development testing approaches that narrow down the source of inward leakage

Finding the source: the role of development testing

A single leakage result shows how much leakage occurred, not where it came from. Because so many factors combine, pinpointing the source usually means changing one variable at a time and comparing results.

Development and pre-approval testing creates room to do this before a device goes for formal approval. Examples include:

  • Trying alternative face seal shapes or materials.
  • Taping or sealing individual joints, such as a visor seal, speech diaphragm or valve housing, to see whether leakage changes.
  • Comparing exhalation valves of different thickness or design.
  • Comparing IL and TIL results to separate filter penetration from other leakage paths.
  • Testing powered devices at different flow rates.
  • Reviewing results exercise by exercise to see which movements affect a particular design.

 

GCS carries out inward leakage and total inward leakage testing to EN 13274-1 for development, pre-approval and approval projects. Get in touch to discuss development testing for your device.

 

 

Frequently asked questions

What is the difference between inward leakage and total inward leakage?

Inward leakage (IL) covers all leakage into the breathing zone except through the filter. Total inward leakage (TIL) also includes what passes through the filter.

Which test agent is used for inward leakage testing?

EN 13274-1 uses either sodium chloride aerosol or sulfur hexafluoride gas. The choice depends on the device type, whether its materials are permeable, and whether IL or TIL is being measured.

At what flow rate are PAPRs and airline devices tested?

Powered and airline devices are typically tested at the manufacturer’s minimum design flow rate. The relevant device standard sets the exact conditions.

Can testing show where a device is leaking?

A single result shows how much leakage occurred, not its source. Comparing results across controlled changes to the device, as in development testing, can narrow it down.